Controller / User Interface | 4-39
CAL Mode – F6 Commands
Keystroke Term
Definition
F6 •
—
—
F6 –
—
—
F6 0
IZero
z
The measured current (mA) in the current output loop when the output is set to zero-scale
during the analyzer Zero, where ‘
z
’ = Output Channels 1..4. Refer to the Output Signal
Assignment code for the outputs assigned to these channels (press
F2 6
from CFG mode).
The actual mA value is displayed next to the corresponding output channel when 4 mA
(Zero) is activated.
1 = IZero1
2 = IZero2
3 = IZero3
4 = IZero4
F6 1 z
ISpan
z
The measured full-scale current (mA) in the current output loop when output is set to
full-scale during the analyzer Span, where ‘
z
’ = Output Channels 1..4. Refer to the Output
Signal Assignment code for the outputs assigned to these channels (press
F2 6
from CFG
mode). The actual mA value is displayed next to the corresponding output channel when
20 mA (Span) is activated.
1 = ISpan1
2 = ISpan2
3 = ISpan3
4 = ISpan4
F6 2
ZCD
z
The Zero Cal Drift, where ‘
z
’ = 1..7. After an Auto-Cal, the Zero Drift of all calculated results
since the last Auto-Cal are recorded. The drift result is displayed in % or PPM.
1 = ZCD1
5 = ZCD5
2 = ZCD2
6 = ZCD6
3 = ZCD3
7 = ZCD7
4 = ZCD4
F6 3
SCD
z
The Span Cal Drift, where ‘
z
’ = 1..7. After an Auto-Cal, the Span Drift of the calculated
result since the last Auto-Cal is recorded. The drift result is displayed in % or PPM.
1 = SCD1
5 = SCD5
2 = SCD2
6 = SCD6
3 = SCD3
7 = SCD7
4 = SCD4
F6 4
O2CD
z
The O
2
Cal Drift, where ‘
z
’ = 1..2. After an Auto-Cal, the O
2
signal drift of the calculated
result since the last Auto-Cal is recorded. The signal drift result is displayed in PPM or %.
(O
2
Sensor is optional; General Purpose applications/analyzers only.)
1 = O2CD1 (O
2
Zero)
2 = O2CD2 (O
2
Span)
F6 5
O2ZVolt
The O
2
Sensor voltage output offset at Zero (air point), determined by the software when
the O
2
Sensor is Zero calibrated.
(O
2
Sensor is optional; General Purpose applications/
analyzers only.)
F6 6
O2SFactor
The O
2
Sensor Span Factor, determined by the software when the O
2
Sensor is Span
calibrated.
(O
2
Sensor is optional; General Purpose applications/analyzers only.)
F6 7
—
—
F6 8
—
—
F6 9
—
—
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...