Maintenance and Troubleshooting
|
5-23
Example:
If replacing Lamp 1, and the filter location with the highest lamp pulse
current-control value is Filter 3, the PMT signal for Filter 3 is displayed
by pressing
F6 1 3
.
If either the
Measure
or
Reference
PMT signal becomes equal to or
greater than 10.000 V at any time during lamp alignment, decrease
the PMT level (CFG
F1 4
to view
PmtLvl
screen) in one-volt incre-
ments until it reaches a stable level. Allow the signals to stabilize and
continue.
13. Slightly loosen the Lamp Clamping Screw. Loosen the Lamp Retaining
Screw and rotate the Lamp Retaining Bracket 90 degrees.
14. Slowly rotate the new lamp to obtain the maximum signal from the
Measure
PMT. Since the User Interface is updated at one-second inter-
vals, use slow, small movements, pausing each turn to view the new
value.
Because the Automatic Lamp Control (
Alc Enable
function) is turned
Off, the displayed signal may not be perfectly steady.
15. Loosen the Locking Screw between the two lamps.
16. To obtain the maximum signal from the
Measure
PMT, adjust the Lamp
Adjusting Screw for the lamp being replaced.
17. To replace the second lamp repeat Steps 9–16.
18. Tighten the Locking Screw located between the two source lamps.
19. Return the Lamp Retaining Bracket to its original position and tighten
the Lamp Retaining Screw. Tighten the Lamp Clamping Screw to se-
cure the lamps.
Do not overtighten these screws; the springs should
not be collapsed.
20. Initiate another Auto-Setup and continue with this procedure only af-
ter the completion number is between 0.75–1.25. See “The Auto-Setup
Sequence” in this chapter.
NOTE
!
CAUTION
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...