Maintenance and Troubleshooting
|
5-61
HS Code
HStatus Alarm Type Condition / Description and Suggested Corrective Action
2
2
Warning
w Out of Range
The concentration of the output exceeds its normal operating full-
scale range by more than 10 %.
Corrective Action:
• From the
Output
screen, check that the full-scale range is
correct for the current sample. If this Warning persists during
normal operation, contact AMETEK. A range change and recali-
bration may be required.
4
3
Fault
f Comm Fault
This serious system Fault condition indicates the
Host Controller
board cannot establish reliable communications with the
Micro-
controller
board.
When this Fault is detected, “
CommFault
” is displayed immedi-
ately on the lower line of the User Interface to indicate the same
serious system alarm as “
f CommFault
”.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclo-
sure, and:
• Check the flat cable between the
Host Controller
board (on
door inside Electronics Enclosure) and
Micro-Interface
board
(mounted to Electronics Enclosure backpan) for proper connec-
tions and inspect it for damage (cuts, nicks, burn marks, etc.).
• Ensure the actuators on switch
S100
on the
Micro-Interface
board are positioned toward the middle of the board. See Figure
5-11 for the location of this switch.
8
4
Fault
f RxData Fault
This serious system Fault condition indicates the
Host Control-
ler
board is not receiving analytical data from the
Microcontroller
board.
When this Fault is detected, “
NoData
” is displayed immediately on
the lower line of the User Interface to indicate the same serious
system alarm as “
f RxData Fault
”.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclo-
sure, and:
• Check the Chopper Wheel Optocoupler cable connection.
• Check the flat cable between the
Optical Bench
board (
J103
)
and the PMT Buffer for proper connections and inspect it for
damage (cuts, nicks, burn marks, etc.).
• Check the PMT high-voltage connections at the
Optical Bench
board (
J101, J102
) for proper connections.
• Check the flat cable between the
Optical Bench
board and the
Micro-Interface
board for proper connections and inspect it for
damage (cuts, nicks, burn marks, etc.).
• Check the Chopper Wheel motor connection.
(Output) RUN
F2 6
1..4
See Ribbon Cable
Interconnect drawing in
Appendix A.
See Ribbon Cable
Interconnect drawing in
Appendix A.
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...