3-46 | Model 920 Multi-Gas Analyzer
3. Record these signals in a log book every month.
It is important to record these values
during
a Manual Zero to
simulate the same conditions as during the original recordings. If the
values are recorded using the sample gas, the results may be skewed
because the composition of the gas can vary. All signals should return
to within 5 % of the values recorded the previous month.
Recording Initial Sample Response Time
The analyzer’s sample response time can be used for two functions:
• To help you set the sample gas flow rate.
• To help you determine if the typical response time is increasing, which
can indicate a plug in the sample system.
Typically, a good response time is approximately 30 seconds when the
sample system is clean and operating at peak efficiency (may vary due to
Sample Line length).
To check and record the initial sample response time:
1. Manually Zero the analyzer.
2. After the Zero is complete, view the RUN mode normal display
and observe the output concentration changes. Observe and record
the time it takes the analyzer to display the first reading after the
analyzer’s
Flow Control
setting is switched from
Continuous Zero
Gas
mode – either “
10
” (
probe tip
) or “1
10
” (
manual
) – to
Analyzer
Control
(“
0
”).
It is important to record the response time only
after
performing a
Manual Zero to simulate the same conditions during the original
recordings.
3. If the sample response time is adequate (first reading after a Zero and
after changing the
Flow Control
mode from “
1
” to “
0
” is approximate-
ly 30 seconds), no further adjustment is necessary.
The sample response time may vary, depending on Sample Line
length.
4. Record these signals in a log book every month.
NOTE
NOTE
See “Manually Zeroing
the Analyzer” in this chapter.
NOTE
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...