Controller / User Interface | 4-37
CAL Mode – F5 Commands
Keystroke Term
Definition
F5 •
—
—
F5 –
—
—
F5 0
Sol:R
z n
Assigns SSR output ‘
n
’ (where ‘
n
’ = 0..2) controlling a Calibration gas solenoid valve to the
calculated result (R), ‘
z
’, where ‘
z
’ = 1..6. The same SSR output can be assigned to more
than one calculated result. This allows a multi-component Calibration gas to be used. If a
calibration gas is not used with one of the calculated results, ‘
n
’ must be set to ‘
0
’.
1 = Sol:R1
4 = Sol:R4
2 = Sol:R2
5 = Sol:R5
3 = Sol:R3
6 = Sol:R6
and where ‘
n
’ = 0..2:
0 = Auto-Zero
1 = Auto-Span1
2 = Auto-Span2
F5 1
Scale
z
The full-scale range for each of the four outputs , where ‘
z
’ = 1..4. If an output is not used,
its full-scale range should be set to ‘
0
’. For example, if Scale 1 = 1000 PPM, Output 1 will
read 20 mA at 1000 PPM and 4 mA at 0 PPM. These outputs are application-specific.
1 = Scale 1
2 = Scale 2
3 = Scale 3
4 = Scale 4
F5 2
FlowCtrl
n
The analyzer system Sample Flow Control mode, where ‘
n
’ = 0..2, 10..12, 100..102, or
110..112.
These codes apply to sample systems using an Aspirator, where ‘
n
’ = 0..2:
0 = Analyzer Control mode (automatic control by analyzer)
1 = Continuous Backpurge mode (manually set by operator)
2 = Continuous Sample mode (manually set by operator)
10 = Continuous Zero Gas Flow mode (manually set by operator) (probe tip)
11 = Continuous Calibration Gas #1 Flow mode (manually set by operator) (probe tip)
12 = Continuous Calibration Gas #2 Flow mode (manually set by operator) (probe tip)
These codes apply to sample systems using a Sample Block Valve:
100 = Analyzer Control mode (automatic control by analyzer)
101 = Continuous Sample Block-Off mode
102 = Continuous Sample mode
110 = Continuous Zero Gas Flow mode (local)
111 = Continuous Calibration Gas #1 Flow mode (local)
112 = Continuous Calibration Gas #2 Flow mode (local)
F5 3
AdjDisable
z
The Calibration Adjustment Disable function can be set to allow (or prevent) automatic ad-
justment of the analyzer Zero offset transmittances and Span factors for all Zero and Span
operating modes (automatic or manual calibration).
To allow automatic adjustments of the analyzer Zero offset and calibration, set this function
to ‘
0
’.
To prevent automatic adjustments of the analyzer Zero offset and calibration, set this func
-
tion to ‘
1
’.
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...