5-6 | Model 920 Multi-Gas Analyzer
Expo Technologies MiniPurge
®
System With eTimer (Optional)
Preventive Maintenance Schedule
Frequency Task
—
If your analyzer uses an Expo Technologies MiniPurge
®
System With
eTimer, refer to the system’s manual for maintenance frequency and
procedures. This manual is shipped with the analyzer.
—
To test the pressure of the Electronics Enclosure, use the Cabinet
Pressure Port on the Solenoid Block.
Every 3 Years
Battery Pack, Expo MiniPurge
®
System (if analyzer is equipped)
Replace the Battery Pack and perform the commissioning tests as
described in the Expo
MiniPurge
®
Type X / ET Size 1 Manual (ML 422)
.
See “Expo Technologies MiniPurge
®
System With eTimer Spare Parts”
in Chapter 6 for spare parts ordering information.
Preventing, Detecting, and Locating a Plug in the
Sample System
Preventing a plug in the sample system:
While various factors can contribute to plugs in the sample system, plugs
can be avoided in many cases by:
• Maintaining the ambient temperature around the analyzer, and the
operating temperatures for each analyzer temperature zone.
From the
MAI
screen, check the real-time temperature for each
Temperature Zone and compare it to its
Set Point
temperature (on the
TStPt
screen).
Also, verify the ambient temperature surrounding the analyzer and
the operating temperature of its sample system is at least 10 °C (18
°F) above the process temperature. The exception to this is the SCU
(optional), whose operating temperature operates nominally 5–15 °C
(41–59 °F) below the sample system temperature. The Sample Probe
(or SCU) and Vent Valve (if used) assemblies should be insulated using
the insulation supplied by AMETEK, plus additional insulation (by
customer) if necessary. It is extremely important to maintain a rela-
tively stable ambient temperature in the vicinity of the analyzer and
its sample system, with no rapid temperature fluctuations.
The risk of developing a collection of liquids in low spots or plugging
in the sample system is further reduced by sloping the Sample Line
and Vent Line (if used) from the analyzer back to the sample extrac-
tion point.
(MAI) RUN
F6 8
1..8
(TStPt) RUN
F4 0
1..4
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...