Installation and Start-Up | 3-33
Purged Analyzers (Hazardous Locations):
Insert the key into the Purge Bypass Switch and switch it to the
“
BYPASS
” position (follow company policy).
[Special Conditions for Safe Use]
The analyzer may only be energized by using the Purge Bypass
Switch with permission of the works manager or his proxy. The per-
mission may only be given when it is made sure that during the time
the system is energized by using this switch an explosive atmosphere
is not present or when the necessary protective measures against
explosion hazard have been taken (“hot permit”).
The analyzer enclosure may not be opened when an explosive atmo-
sphere is present.
Close the Analyzer fuse.
Apply main AC power to the analyzer by closing the explosion-proof
power-disconnect switch.
2. Immediately following power-up, the
Host Controller
board software
version number will appear on the top line of the User Interface and
the message “
Reset in Progress
” will appear on the bottom line.
If the “
Reset in Progress
” message does not appear within approxi-
mately 2 seconds, simultaneously press
•
and
Ent.
If the “
Reset in Progress
” message still does not appear, take all neces-
sary safety precautions to power down the analyzer, open all of the
fuses for the temperature zone circuitry, and check the AC wiring to
ensure it is properly connected.
After powering down the analyzer, wait 5 minutes to allow the high-
voltage capacitors in the source-lamp power supply to discharge.
Review the Analyzer (AC) Power Connections drawing(s) and check
the wiring terminations from the power supply source to the analyzer.
Take all necessary safety precautions and power up the analyzer
(Step 1). Check the User Interface again to see if the message appears.
If it does, continue with the next step.
!
WARNING
!
WARNING
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...