Maintenance and Troubleshooting
|
5-29
Manipulating the Completion Number, PMT Level, and PMT Balance
In certain situations (for example, degradation of PMTs or lamps, or the
optics are contaminated), it may be necessary to change the
P300
and
P301
jumper settings on the
Optical Bench
board. Changing the jumper posi-
tions provides coarse gain adjustment of the
Reference
PMT and therefore
changes the
completion number
result,
PMT Level,
and
PMT Balance
.
There are six jumper positions (four in
P300
and two in
P301
) – at least
one jumper must be in one of the six positions at all times. The jumper
combinations control the coarse supply voltage ranges of the two PMTs,
Reference
-path and
Measure
-path. The Optical Bench Auto-Setup proce-
dure performs fine adjustment to the PMT supply voltages at one jumper
combination. The jumper combinations affect the Auto-Setup in terms of
reducing the
completion number
, increasing or decreasing the resulting
PMT Balance and, to a lesser degree, the PMT Level.
Figure 5-3 describes the effect on the PMT Balance and PMT Gain in rela-
tion to different jumper positions.
To achieve an acceptable
completion number
,
PMT Balance
, and
PMT
Level
:
1. If jumpers have been reconfigured, perform an
Auto-Setup
.
If the PMT signals are outside their normal range of operation after
the Auto-Setup is complete (typical range is 4.5–7.8 VDC), adjust the
Lamp Max
setting slightly and start another Auto-Setup.
2. After an acceptable
completion number
,
PMT Balance
, and
PMT
Level
has been achieved, perform a Manual Zero.
3. After the Zero is complete, change the
Flow Control
setting back to
Analyzer Control
mode.
(AutoSetup) CFG
F1
•
See “Manually Zeroing
the Analyzer” in Chapter 3.
NOTE
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...