5-60 | Model 920 Multi-Gas Analyzer
Host Controller Board Alarm Conditions and
Corrective Action
This section lists valid alarms (errors) that originate from the
Host
Controller
board, including the
Alarm Type
(Fault or Warning), name and
reason the alarm is set (
Condition/Description
), and
Suggested Corrective
Action
for each. The Host Controller Status Codes (
HS Code
) and Host
Controller numbers (
HStatus
) to press on the keypad to view the current
status or alarms are also listed.
The possible Host Controller board alarms are polled individually with the
keypad command
F5 4
z
, where ‘
z
’ (
z
=
1..7
) is defined by
HStatus
. For each
‘
z
’, the corresponding alarm message description is displayed if the alarm
exists; otherwise the message “
HS
z
OK
” appears.
On the User Interface, the ‘w’ or ‘f’ preceding an alarm indicates a
‘w’arning or ‘f’ault condition.
HS Code
HStatus Alarm Type Condition / Description and Suggested Corrective Action
0
—
—
HSz OK
If the HS Code returned is ‘
0
’ and the message displayed is “
HS
z
OK
”, it indicates that the Warning or Fault being viewed does not
exist (where ‘
z
’ is the HStatus number 1..7 assigned to that alarm).
This is not an alarm condition.
Corrective Action:
No action is required.
1
1
Warning
w EEPROM Full
The EEPROM (nonvolatile memory) has exceeded the safe num-
ber of times that configuration data has been rewritten to it (95 %
used).
Corrective Action:
• Replace the EEPROM as soon as possible.
Contact AMETEK
to verify operation before removing the EEPROM
and for
information about installing the new EEPROM to ensure your
analyzer will operate the same as it did prior to replacing the
EEPROM.
NOTE
(HStatus) RUN
F5 4
1..7
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...