Controller / User Interface | 4-35
CAL Mode – F3 Commands
Keystroke Term
Definition
F3 •
—
—
F3 –
—
—
F3 y
Timery
Sets the duration the solenoid valve controlling the Zero or Calibration gas connected to
SSR output ‘y’ is energized during the Auto-Zero, Auto-Span, and Auto-Calibration func-
tions, where ‘y’ =
F3 0..F3 2
.
F3 0 = Timer0
F3 1 = Timer1
F3 2 = Timer2 –
Used only with O
2
Sensor (General Purpose applications)
Timer0
is the Zero solenoid timer, while
Timer1
and
Timer2
are Calibration gas mixture
solenoid timers. The duration can be set from 0–255 minutes in one-minute increments.
Setting the timer to ‘
0
’ turns it off.
F3 3
PrbTz
The
Probe Blow-Back Timers
allow you to set up the Blow-Back feature for systems using
a sample probe with a Blow-Back function (optional), where ‘z’ = 1..3.
1 = PrbT1: The Probe Blow-Back Interval timer allows you to enter the interval-
time between Blow-Backs (minutes). When an interval-time is entered, it is
displayed along with a countdown timer (time remaining until the next Blow-Back).
2 = PrbT2: The Probe Blow-Back Duration timer allows you to enter the duration
of each Blow-Back (seconds).
3 = PrbT3: The Probe Blow-Back Pulse Width timer allows you to enter the duration
(in seconds) the analyzer will allow the Blow-Back system to build up pressure
between pulses (air bursts, which blow back debris into the stack).
F3 4
—
—
F3 5
—
—
F3 6
IntTime
Sets the duration (seconds) over which the reading for each Manual Zero or Span is aver-
aged. Setting the timer to ‘
0
’ turns the function off.
F3 7
SDelay
Sets the time delay for the
CAL
relay to continue indicating that a calibration is in progress
after the calibration is complete. This permits a smooth transition from the Zero/Span gas
back to the sample gas. The delay can be set from 0–255 minutes in one-minute incre-
ments. Setting the delay to ‘
0
’ turns the function off.
The delay also occurs when exiting CAL or CFG mode.
F3 8
AZInt
Sets the time interval (hours) between Auto-Zeroes. This allows Auto-Zero to be performed
more frequently than Auto-Calibration. The interval can be set from 0–999 hours in one-
hour increments. Setting this interval to ‘
0
’ turns the function off.
The Auto-Zero function will be included in the Auto-Calibration sequence if Auto-Zero is
enabled and
Timer0
duration is not ‘
0
’.
Also displays the time remaining until the next Auto-Zero (hours and minutes).
F3 9
ACal
The Auto-Calibration Interval setting sets the duration (hours) between Auto-Calibrations.
The interval can be set from 0–999 hours in one-hour increments. Setting this function to ‘
0
’
turns it off.
Also displays the time remaining until the next timed Auto-Calibration (hours and minutes).
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...