41-621.1 HART
®
to Modbus
®
Adaptor Operating Manual
121
Error
Code
Model 706
Model JM4
Model R96
1
Software Error
SW Error (Main)
Software Error
2
RAM Error
RAM Error (Main)
RAM Error
3
ADC Failure
ADC Error (Main)
ADC Failure
4
EEPROM Error
EEPROM Error
EEPROM Error
5
Analog Board Error
CoP in Flash Mode
Analog Board Error
6
Analog Output Error
SW Conflict (CoP)
Analog Output Error
7
Spare 1
Spare 1
Spare 1
8
Default Parameters
Analog Board Error
Default Parameters
9
No Probe
SW Error (CoP)
No Antenna
10
No Fiducial
RAM Error (CoP)
Spare 2
11
No Echoes
ADC Error (CoP)
No Fiducial
12
Upper Echo Lost
Spare 2
Too Many Echoes
13
Spare 2
Analog Ouput Error
Safety Zone Alarm
14
EoP > Probe End
No Probe
Echo Lost
15
Level Below Probe End
Probe Memory Error
Spare Indicator 3
16
EoP Below Probe End
Probe Info Corrupt
Configuration Conflict
17
Safety Zone Alarm
Spare 3
High Volume Alarm
18
Config Conflict
New Probe
Spare Indicator 4
19
Hi Volume Alarm
Default Parameters
Initializing
20
Hi Flow Alarm
No Float Detected
Configuration Changed
21
Spare 3
Spare 4
Spare Indicator 5
22
Initializing
Config Conflict
Ramp Slope Error
23
Analog Output Fixed
Hi Volume Alarm
High Electronics Temp
24
Config Changed
Spare 5
Low Electronics Temp
25
Spare 4
Extra Float Detected
Calibration Required
26
Spare 5
2nd Float Missing
Echo Rejection Invalid
27
Spare 6
Initializing
Spare Indicator 6
28
Ramp Interval Error
Config Changed
Inferred Level
29
Hi Elec Temp
Spare 6
Adjust Analog Output
30
Lo Elec Temp
Xmtr Calib Req'd
Low Supply Voltage
31
Calib Req'd
Spare 7
Spare Indicator 7
32
Echo Rej Invalid
Temp Calib Req'd
Spare Indicator 8
33
Spare 7
Hi Elec Temp
Marginal Echo
34
Inferred Level
Lo Elec Temp
Hi Surface Velocity
35
Adj Analog Output
Spare 8
Spare Indicator 9
36
Totalizer Data Lost
Spare 9
Spare Indicator 10
37
No Probe Target
Adj Analog Output
Sequence Record
38
Low Supply Voltage
Low Supply Voltage
39
Dry Probe
Spare 10
40
Spare 8
Lo Echo Strength
41
Lo Echo Strength
Lo Ifc Echo Strength
42
Lo Ifc Echo Strength
Hi Noise / LvlThresh
43
Spare 9
Hi Noise / IfcThresh
44
Spare 10
Spare 10
45
Sequence Record
Sequence Record
Note: Diagnostics mapped to the NE 107 Failure category will appear as a LM Error, those mapped to other