80
41-621.1 HART
®
to Modbus
®
Adaptor Operating Manual
17.5.24
Select Holding Registers from the tree-view window of the PCCU32.
17.5.25
Select the Blocking Distance tab.
17.5.26
Change the Value to the desired distance.
17.5.27
Click on Send at the bottom of the window.
17.5.28
To check if the value has been accepted, follow the steps in section 17.4 changing
selections as needed to create a request block to read the Blocking Distance parameter
from the HMA or attached HART device.
Summary of Contents for Jupiter JM4
Page 2: ......