Proline Edition X cooling thermostats
94
/ October 4, 2018
Command
Explanation
START
Switches the unit on (after Standby).
See safety information
(
STOP
Switches the unit into Standby (pump, heater, cooling unit OFF).
RMP_SELECT_X
Selection of the program (1
— 5) to which the further instructions apply.
When the unit is switched on, program 5 is selected automatically.
RMP_START
Start the programmer.
RMP_PAUSE
Hold (pause) the programmer.
RMP_CONT
Restart the programmer after pause.
RMP_STOP
Terminate the program.
RMP_RESET
Delete the program (all Segments).
RMP_OUT_00_XXX.XX_
XXXXX_XXX.XX_X
Set a programmer segment (temperature, time, tolerance and pump level).
A segment is added and appropriate values are applied to it.
RMP_OUT_02_XXX
Number of times the program runs: 0 = unlimited / 1
— 250.
For ”_“ use also ” ” (blank character).
Response from thermostat ”OK“ or in case of error ” ERR_X“ (RS-485 interface
e.g. “A015_OK” or in case of error ”A015_ERR_X”.).
Permitted data formats:
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
8.3.6 Read commands (Data requested from the thermostat)
Command
Explanation
IN_PV_00
Read bath temperature (outflow temperature).
IN_PV_01
Indication of the controlled temperature (internal/ external Pt /
exteral analogue / external serial).
IN_PV_03
Read external temperature TE (Pt100).
IN_PV_04
Read external temperature TE (analogue input).
IN_PV_05
Read current bath level.
IN_PV_10
Read bath temperature (outflow temperature)
in 0.001 °C.
IN_PV_13
Read external temperature TE (Pt100)
in 0.001 °C.
IN_SP_00
Read temperature setpoint.
IN_SP_01
Read current pump power level
IN_SP_02
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
IN_SP_03
Read current overtemperature switch-off point.
IN_SP_04
Read current outflow temperature limit TiH.
IN_SP_05
Read current outflow temperature limit TiL.
IN_PAR_00
Read current value of Xp.
IN_PAR_01
Read current value of Tn (181 = OFF).
IN_PAR_02
Read current value of Tv.
IN_PAR_03
Read current value of Td.
IN_PAR_04
Read current value of KpE.
IN_PAR_05
Read current value of TnE (980 = OFF).