Proline Edition X cooling thermostats
64
YACE0100 / October 4, 2018
Using the cursor keys move the black background to the field, which you would like to change.
It can be edited by pressing the key
(see following pages).
The soft key
Insert inserts in the marked line a new segment that has a default value
taken from the previous segment with the exception of the Tolerance field. The Tolerance is
always specified as 0.00. All following segment lines will be moved one line downwards.
In the above window Segment 1 was created in this way.
Continue with
to the fields
"Time
"Tolerance". See program example in (
If there is no entry in the "Time" field, the bath temperature is approached as quickly as
possible. With a time entry the final temperature is obtained exactly after the time expires
(ramp).
The entry in the field "Tolerance" field defines how accurately the final temperature is to be
obtained before the next segment is processed.
If the tolerance range has been selected too small, it may be that the program does not
continue, because the required tolerance is never achieved.
External temperature control: Especially with ramps, a too close tolerance range can
cause undesired delays in the start phase of the ramp.
No.
Pump
Out 1
Out 2
Out 3
Start
-------
------- -------
-------
1
4
------- -------
-------
Pump
Menu
End
Insert Delete
Then continue with
to the pump and
signal output setting.
The right-hand part of the entry table
appears as shown on the left.
Here, in the "Pump" field, the pump level
and, in the fields "Out 1" to "Out 3", the
contact outputs of the contact mode
(accessory) can be programmed. With
the setting "-------" the starting value is
retained which was either set before the
program start or was defined by a
previous segment in the running
program. Further details are given on the
following pages.