Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
Technical data
109
RP 845C RP 855C RP 870C RP 890C RP1290C RP1840C RP1845C RP3530C
Overall dims. B x L
mm
285 x 430
400 x 540
375 x 540
495 x 615
495 x 615
375 x 465
400 x 540
375 x 615
Overall dim. H
mm
688
770
735
735
735
688
770
740
Weight
kg
41
60
68
100
100
46
61
51
Power consumption
230 V; 50 Hz
115 V; 60 Hz
200 V; 50/60 Hz
115 V; 50/60 Hz
208-220 V; 60 Hz
kW
kW
kW
kW
kW
3.6
1.8
3.2
1.6
3.6
IP protection class
IP 2 1
Safety equipment
Class III to DIN 12876-1, FL suitable for flammable and non-flammable liquids
Class of protection for
electrical operating equipment
Protection class I according to DIN EN 61140; VDE 0140-1
The cooling power will be reduced by 30
– 35 W, when pump level 8 is chosen instead of level 3.
Put-on Command remote control: 56 mm higher.
Mains connection data
Mains connection
data
RP 845 C RP 855 C RP 870 C RP 890 C RP 1290C RP 1840C RP 1845C RP 3530C
230 V ±10 %; 50 Hz
X
X
X
X
X
X
X
X
115 V ±10 %; 60 Hz
X
-----
-----
-----
-----
X
-----
X
200 V ±10 %; 50/60 Hz
X
X
X
X
X
X
X
X
100 V ±10 %; 50/60 Hz
X
-----
-----
-----
-----
X
-----
X
208-220 V ±10 %;
60 Hz
X
X
X
X
X
X
X
X
Technical modifications reserved.