Proline Edition X cooling thermostats
54
YACE0100 / October 4, 2018
The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with
or End ,
or quit the window with
without making
changes.
7.8.8 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
dEF 1
3x
to
MNEnu
2x
to
CAL..
1x
Call offset works setting for the internal temperature probe
dEF
1
.
dEF 1
is displayed.
3 sec
The works setting is accepted after pressing for 3 seconds and
donE
is displayed.
Command
Default
Open the device parameter menu via
the soft key Menu .
With the cursor keys continue to
Settings
Calibration
intern Pt100
Default
.
The window shown adjacent appears.
Confirm selection with
.
dEF 1
°C
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temperature value of the ref.
temp. measurement device:
20.
15
Min: -50.0°C Max: 302.00
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default