Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
79
With the menu
Setpoint
the set
temperature for the test run can be set.
The bath temperature will oscillate less
than about ±15 Kelvin around the set
temperature.
Change the display in the adjacent
window and accept with
.
With the menu
Identification
the optimal
control parameters for internal control or
for the internal control and the external
control can be detected automatically.
To detect the control parameters for the
external application, a temperature probe
must be connected to the thermostat.
With the menu
Actual Parameters
the
actual set control parameters can be
stored in parameter set 9.
After the test run the detected control
parameters will be taken over as control
parameters automatically. If the para-
meters found do not fulfil your
expectations, the before set parameters
can be restored (
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
30,00°C
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
intern
extern
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
Store in Set 9