Proline Edition X cooling thermostats
108
Technical data
October 4, 2018
11
Technical data and diagrams
The figures have been determined according to DIN 12876.
RP 845C RP 855C RP 870C RP 890C RP1290C RP1840C RP1845C RP3530C
Operating temp.- ACC range
°C
-45
– 200
-55
– 200
-70
– 200
-90
– 200
-88
– 200
-40
– 200
-50
– 200
-35
– 200
Ambient temp. range
°C
5
– 40
Relative humidity
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to 50 % relative
humidity at 40 °C
Storage temperature range
°C
-20
– 44
Setting resolution
°C
0.1 / 0.01 (Master); 0.01 (Command)
Display resolution
°C
Master: 0.01 Command: 0,1 / 0,01 / 0,001
Accuracy of indication
±0.2 K can be calibrated additively (
1.2 )
Temperature stability
K
±0.01
±0.01
±0.02
±0.02
±0.02
±0.01
±0.01
±0.02
Heater power
230 V; 50 Hz
115 V; 60 Hz
200 V; 50/60 Hz
100 V; 50/60 Hz
208-220 V; 60 Hz
kW
kW
kW
kW
kW
3.5
1.75
2.7
1.3
2.9
C
o
o
lin
g
p
o
w
e
r
a
t
2
0
°C
t
a
mb
(
P
u
mp
L
e
v
e
l
3
)
@b
a
th
t
e
mp
e
ra
tu
re
with thermal
transfer oil
200 °C
kW
1.0
1.7
0.5
0.5
0.5
1.0
1.7
1.0
with ethanol
20 °C
kW
0.8
1.6
0.38
1.1
1.1
0.9
1.6
0.9
0 °C
kW
0.7
1.1
0.36
1.0
1.0
0.7
1.1
0.7
-20 °C
kW
0.36
0.6
0.33
0.9
0.9
0.35
0.55
0.3
-30 °C
kW
0.22
0.38
0.3
0.83
0.83
0.2
0.32
0.15
-40 °C
kW
0.11
0.21
0.25
0.75
0.75
0.09
0.18
--
-45 °C
kW
0.05
0.15
--
0.1
--
-50 °C
kW
--
0.1
0.25
0.58
0.58
--
0.045
--
-55 °C
kW
--
0.04
--
--
--
--
--
--
-60 °C
kW
--
--
0.2
0.42
0.42
--
--
--
-70 °C
kW
--
--
0.1
0.24
0.24
--
--
--
-80 °C
kW
--
--
--
0.13
0.13
--
--
--
-88 °C
kW
--
--
--
0.04
0.04
--
--
--
Pump type
Pressure/suction pump, 8 power levels
Discharge pressure max.
bar
0.7 at Pump Power Level 8
Intake suction max.
bar
0.4 at Pump Power Level 8
Flow rate max. (pressure)
L/min
25 at Pump Power Level 8
Flow rate max. (suction)
L/min
23 at Pump Power Level 8
Hose connections
Thread M16 x 1; olives 13 mm external diameter
Bath volume from
– to
L
5.5
– 8
7.4
– 8.9
14.8
– 17.8
12.5
– 19
23
– 35
Bath opening B x L
mm
150 x 150
300 x 150
300 x 200
300 x 350
Bath depth / usable depth
mm
200 / 180
250 / 230
Height to top of bath
mm
488
570
535
535
535
488
570
540