Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
85
7.14.5 Pump-motor supervision: Overload or blockage
The SelfCheck Assistant monitors the Varioflex pump:
1. Alarm sounds as dual-tone signal for pump-motor overload or
blockage.
2. Display of
bL0C
signals blockage.
3. The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Pump and refrigerating unit are switched off electronically.
Find the cause of the fault. Perhaps the viscosity of the heat
transfer liquid is too high or the pump is blocked.
Press the Enter key.
Also press this key if the unit has been switched off in the fault
state.
Command
Pump-motor alarm!
Pump-motor alarm is shown in the display and signifies that
unlocking is only possible on the Master control panel.
7.14.6 Pump-motor supervision: Dry running
The SelfCheck Assistant monitors the Varioflex pump:
1. Alarm sounds as dual-tone signal when the pump runs without
liquid. This can only occur when the float level measurement has
failed.
2. The display of
PuLEU
signals that the SelfCheck Assistant
has detected a pump low level.
3. The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Pump and refrigerating unit are switched off electronically.
The cause of the failure of the level measurement with the floatation sensor must be
found and rectified. Perhaps it is blocked by foreign bodies in the bath.
Press the Enter key.
Also press this key if the unit has been switched off in the fault
state.
Command
Alarm! Low level (pump)
Alarm! Low level (pump) is shown in the display and signifies
that unlocking is only possible on the Master control panel.
Pump alarm
bL0C
°C
Pump alarm
PuLEU
°C
Summary of Contents for Proline RP 1290 C
Page 2: ......
Page 4: ......
Page 115: ...Proline Edition X cooling thermostats YACE0100 October 4 2018 115 Blank page ...
Page 116: ...Proline Edition X cooling thermostats 116 YACE0100 October 4 2018 Blank page ...
Page 118: ......
Page 119: ......