Editing the control parameter sets
The change in the control parameters is explained in Section 7.13.1/7.13.2 (internal / external). Once
the value has been changed and confirmed, the set number, e.g.
Set 3
and
Upload actual
, the new
value is accepted into the control parameter set to be changed (Set 3) via the command
Control parameter sets
.
7.13.4 Self Adaption
The function Self Adaption can be used to detect automatically the optimal control parameters for
internal or external control.
The Self Adaptation can be performed only on a device with active cooling.
This function is available from software version 2.18 of Command. For thermostats with an older
software version a software update is necessary.
The Self Adaption determines the parameters by a test run of the thermostat. In this case the
thermostat and, if applicable, the external application must be ready for operation (
The Self Adaption will be performed with the actually set pump step. Best results can be achieved
with high pump steps.
The test run must be performed at a passive system, this means that during the test run a exo- or
endtherrmic reaction
mustn’t take place.
The test run takes depending on the external application about 30 minutes to 3 hours. The bath
temperature will oscillate in this time less than about ±15 Kelvin around the set temperature. After the
test run the detected control parameters will be taken over as control parameters automatically.
Command
Self Adaption
Open the device parameter menu via the
soft key
Menu .
With the cursor keys continue to:
Control
Control Parameters
Self Adaption
Settings
.
Confirm selection with
.
The window shown adjacent appears.
With the menu
Status
the test run of the
Self Adaption can be started. When the
Self Adaption is finished, the test run will
be terminated automatically.
As soon as start
is pressed, in the
sofkey area the information
Adaption on
will be displayed followed by the actual
status of the test run.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Settings
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
Start