Proline Edition X cooling thermostats
48
YACE0100 / October 4, 2018
Enter the desired limit temperature.
Accept the change with
,
or quit the window with
without making
changes.
7.8.4 Setpoint offset operating mode
With this function it is possible to apply an offset value to the temperature provided by the external
temperature probe or a module and then to use it as the setpoint. The bath temperature can, for
example, be operated at -25 °C below the temperature of a reactor, which is being measured by the
external temperature probe.
Master
SEtq.
3x
to
MNEnu
to
PArA
5x
SEtq.
in the following submenu, the relative setpoint (Set)
and the source of the actual value can be entered.
Continue with
.
The offset operating mode is currently deactivated.
To activate it press
and with
change to
EP
(external Pt100),
EA
(external via analog module) or
ES
(external via serial
module).
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with the Enter key.
Continue with
. The left-hand window is displayed.
To activate the relative offset input press
and change the value
with
or
.
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with the Enter key.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Lower limit (T il)
-50.
0
Min: -50.0 °C Max: 301.0 °C
SEtq.
°C
r 0FF
°C
r0FFS
°C
Summary of Contents for Proline RP 1290 C
Page 2: ......
Page 4: ......
Page 115: ...Proline Edition X cooling thermostats YACE0100 October 4 2018 115 Blank page ...
Page 116: ...Proline Edition X cooling thermostats 116 YACE0100 October 4 2018 Blank page ...
Page 118: ......
Page 119: ......