Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
73
7.13.1.1 Proven settings for control parameters and pump (integral measurement probe)
Instrument
Type
Heat transfer liquid
Xp
_P
Tn
tn
Tv
tu
Td
td
Pump Level
RP 845
Water
4.0
50
35
6
4
RP 845
Ethanol
7.0
50
35
6
4
RP 845
Ethanol
5.5
30
21
3.6
4
RP 855
Ethanol
7.0
50
35
6
4
RP 855
Ethanol
7.0
30
---
---
4
RP 855
Water
4.0
30
---
---
4
RP 855
Water-Glycol
4.0
30
---
---
4
RP 1845
Ethanol
5.5
50
40
6
5
Technical changes reserved!
7.13.2 External control variable (External measurement probe)
You only need to read further here if you have connected an external temperature probe or the actual
temperature is read in from a module (and you have activated it as control variable according to Section
7.7.4).
Only modify the control parameters if you have knowledge of control techniques.
The control system for external actual values is implemented for improvement of the control behavior as
a two-stage cascade controller. A "master controller" determines the "internal setpoint", from the
temperature setpoint and the external temperature, passed to the slave controller. The control value of
the slave controller controls the heating and cooling.
When a setpoint step change is specified, it may be that the optimum control would set a bath
temperature, which might significantly exceed the temperature desired on the external vessel. There is
a correction limitation, which specifies the maximum permissible deviation between the temperature on
the external load and the bath temperature.
Master
P1d..
3x
to
MNEnu
1x
to
P1d..
As shown in the menu structure (
7.6), the parameters for the
external control variable can now be set.
Select the parameter with
or
and confirm with
. The set
value is displayed. Adapt it with
or
confirm with
.
Example: Proportional factor of the master controller Kpe = 1.5.
Parameters
Ma
ster controller (PIDT
1
-controller):
Ma
proportional factor:
EP
= Kpe
as factor.
Ma
proportional range:
Eb
= Prop_E in kelvin.
Ma
reset time:
En
= Tne
in seconds.
Ma
derivative time (auto/man)
Ev
= Tve
in seconds.
The thermostat logic system only permits values with Tne > Tve!
Ma
damping time (auto/man)
Ed
= Tde
in seconds.
EP !5
°C