background image

INTRODUCTION

This instrument is a portable 3½ digit, compact-sized
digital thermometer designed to use external K-type
thermocouple  as  temperature  sensor.  Temperature
indication follows National Bureau of Standards and
IEC  584  temperature/voltage  tables  for  K-type
thermocouples.  One  K-type  thermocouple  is  sup-
plied with the thermometer.

SAFETY  INFORMATION

It is recommended that you read the safety and op-
eration instructions before using the thermometer.
Designed to meet IEC-1010-1, CE-EMC.

WARNING

TO AVOID ELECTRICAL SHOCK, DO NOT USE
THIS INSTRUMENT WHEN VOLTAGES AT THE
MEASUREMENT SURFACE EXCEED 24V AC OR
60V DC.

WARNING

TO AVOID DAMAGE OR BURNS, DO NOT MAKE
TEMPERATURE MEASUREMENTS IN MICRO-
WAVE OVENS.

CAUTION

Repeated sharp flexing can break the thermocouple
leads. To prolong lead life, avoid sharp bends in the
leads, especially near the connector.
The 

 symbol on the instrument indicates that the

operator must refer to an explanation in this manual.

SPECIFICATIONS

ELECTRICAL

Temperature Scale:

Celsius or Fahrenheit user-selectable

Measurement Range:

-50°C to 1300°C, ( -58°F to 2000°F )

Resolution: 

1°C or 1°F, 0.1°C or 0.1°F

Accuracy:

Accuracy is specified for operating temperatures over the range
of 18°C to 28°C ( 64°F to 82°F ) , for 1 year,not including
thermocouple error.
±( 2°C ) -50°C to 0°C
±( 0.3% rdg + 1°C ) 0°C to 600°C
±( 0.5% rdg + 1°C ) 600°C to 1300°C
±( 4°F ) -58°F to 32°F
±( 0.3% rdg + 2°F ) 32°F to 1100°F
±( 0.5% rdg + 2°F ) 1100°F to 2000°F

Temperature Coefficient:

0.1x( specified accuracy ) per °C . ( 0°C to 18°C , 28°C to 50°C ).

Input Protection:

60V dc or 24V rms ac maximum input voltage on any combina-
tion of input pins.

Reading Rate:

2.5 times per second.

Input Connector:

Accepts standard miniature therm ocouple connectors ( flat
blades spaced 7.9mm , center to center ) .

ENVIRONMENTAL

Ambient Operation Range:

0°C to 50°C, (32°F to 122°F)

Storage Temperature:

-20°C to 60°C, (-4°F to 140°F)

Relative Humidity:

0% to 80% (0°C to 35°C) (32°F to 95°F)
0% to 70% (35°C to 50°C) (95°F to 122°F)

GENERAL

Display:

3½  digit  liquid  crystal  display  (LCD)  with  maxi-
mum  reading  of  1999

Battery:

Standard 9V battery (NEDA 1604,IEC 6F22)

Battery Life:

100 hours typical with carbon zinc battery
200 hours typical with alkaline battery

Dimensions:

 184mm (H) x 62mm (W) x35mm (D)

Weight:

 10.6oz (300g) including holster

Supplied Probe:

4-foot  type 

K

  thermocouple  bead  probe  (teflon

tape  insulated).  Maximum  insulation  temperature
260°C (500°F). Probe accuracy ±2.2°C or ±0.75%
of reading (whichever is greater) from 0° to 800°C.

OPERATING   INSTRUCTIONS

DIGITAL  THERMOMETER

T1

T2

T1-T2

ON   OFF

C

°

1

°

0.1

°

/

F

°

/

MIN

MAX

/

HOLD

THERMOMETER

OFFSET

T1

T2

Reviews: