Handbook OPTIFLEX 1300 C
81 OPTIFLEX
1300
C
Environment
Ambient temperature:
-40…+80°C/ -40…+175°F; EExi: **
Storage temperature:
-40…+85°C/
-40…+185°F
Accuracy:
0,05%/ rel. 20 mA; 20°C/ 70°F
Resolution:
± 2µA
Temperature drift:
Typically 50 ppm/K
Electromagnetic
compatiblity:
Conforms to Directive 89/336/CEE in conjunction with EN 61326 (1+2) -
immunity and emissions
Process
Temperature
Flange temperature:
-40…+200°C/ -40…+390°F**
Thermal
shock
resistance:
100°C/ min
Pressure
Operating pressure:
-1…40 bar/ -14.5…580 psig**;
subject to process connection used and flange
temperature
Min. dielectric constant
Level in direct mode:
1.4 for coaxial probe,
1.6 for single and double probes
Interface:
ε
r
(interface)
>>
ε
r
(level)
²
Level in TBF mode:
1.1
Protection category:
IP 66/67 equivalent to NEMA 6-6X
Mechanical data
Dimensions:
See section 7.2 Dimensions and weights
Weight:
See section 7.2 Dimensions and weights
Material: Housing
Aluminium
Single rod
Stainless steel (316 L); Hastelloy® C-22
Double rod
Stainless steel (316 L); Hastelloy® C-22
Coaxial
Stainless steel (316 L); Hastelloy® C-22
Single cable
Stainless steel (316);
Hastelloy® C-22 (only cable Ø4 mm / 0.15")
Double cable
Stainless steel (316)
Process fitting
Stainless steel (316 L); Hastelloy® C-22
Gaskets
Viton® (-40…+200°C/ -40…+390°F);
Kalrez® 6375 (-20…+200°C/ -5 …+390°F)
Process connections:
Thread: G 3/4…1 1/2; NPT 3/4…1 1/2; Flanges: DN 25…DN 150
(PN 40/PN 16); 1"…8" (150 lbs/ 300 lbs); 10 K (40…100A)
Electrical connection for 2-wire power supply
Cable entry:
M20x1.5; 1/2 NPT; G ½ (not for FM/CSA devices)
Terminals: 0.5…1.5
mm²
Terminals output 1/HART:
Non-Ex/ EEx i
24 V DC (14…30 V DC)** ***
EEx d
24 V DC (20…36 V DC)** ***
Terminals output 2:
Non-Ex/ EEx i/ EEx d 24 V DC (1030 V DC)** ***
Human-machine interfaces
PACTware:
Open source device type manager for remote display and configuration
Display screen (optional):
160x160 pixels in 8-step grey scale with 4 keys. Wizard-driven.
9 languages available (English and a 2nd language)
* TBF is the abbreviation for Tank Bottom Following. Refer to section 8.2 for further information.
** see supplementary operating instructions or approval certificate
*** respect correct polarity of the electrical system
Summary of Contents for optiplex 1300 c
Page 1: ......
Page 49: ...Handbook OPTIFLEX 1300 C 49 OPTIFLEX 1300 C 1 Volume measurement of liquids...
Page 50: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 50 2 Volume measurement of solids...
Page 51: ...Handbook OPTIFLEX 1300 C 51 OPTIFLEX 1300 C 3 Mass measurement of liquids or solids...
Page 62: ...Handbook OPTIFLEX 1300 C OPTIFLEX 1300 C 62 Finding the advanced setup menu...
Page 95: ...Handbook OPTIFLEX 1300 C 95 OPTIFLEX 1300 C Note...