Product build-up
Product can accumulate under the nozzle: this may weaken the measurement pulse. Avoid cavities that
permit the build-up of deposits.
You can configure the level meter to attenuate the effects of this build-up.
For further information, refer to section 4.2.4: Setup, advanced setup. Follow the procedures for modifying
Gain and voltage amplitude of the level and/or interface thresholds.
For further assistance, contact the supplier.
Cable probes
Installing level meters with single or twin cable probes
1. Support the housing and the probe using two people.
2. Insert the probe by holding the device more than one meter above the opening to avoid over bending the
cable.
Do not over bend probe!
Risk of measurement error!