Ch.6 Troubleshooting
< 131 >
DRYPRO Vstage MODEL 793 Operation Manual Ver.1.01 2005.1
9
Install the Deodorant Filter Case.
- With your left hand holding the handle (green) at the
front of the Deodorant Filter Case, and your right hand
holding the handle (green) at the top, install the
Deodorant Filter Case.
10
Turn the Lever-A clockwise.
- Engages the Deodorant filter Case Lock.
11
Close the Front Top Cover by your hand.
- "READY" will be displayed in the message column
when the jam release sequence is completed, and the
DRYPRO 793 becomes ready to print.
- The print job that has been interrupted due to the film
jam will be reregistered, and printed with a priority.
When failed to install the Deodorant Filter Case or forgot install it, an error message
"E4723 No Filter" will be displayed. Make sure to install the Deodorant Filter Case.
When there were other print jobs left, they will be automatically printed after the filter
installation.
Animation of Jam
Release in the Cooling
Unit.
7
Open the Ejection Transport Unit Cover (1 in the
left fig.), and remove the jam film (2 in the left
fig.).
8
Close the Ejection Transport Unit Cover.
Summary of Contents for DRYPRO Model 793
Page 8: ...About This Manual 6 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 18: ...Ch 1 Safety Precautions Warnings 16 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 28: ...Ch 2 Product Outline 26 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 48: ...Ch 3 Operation 46 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 100: ...Ch 4 Maintenance Mode 98 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 124: ...Ch 5 Web Maintenance Mode 122 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 150: ...Ch 7 Care Maintenance 148 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 154: ......
Page 155: ......