5-3-5 PRINT CONDITION <PRINT SETUP>
Allows to assign a default print condition to each diagnostic device for the case that the
print condition is not designated by the diagnostic device.
Item
Default
Descriptions
LUT SELECT
1
Select the LUT for the print condition.
Select from "1" - "7".
Use "LUT Setup" of the Maintenance Mode to set the details of LUT.
DENSITY
0
Sets the density of the image. Select from "-7" - "+7". More negative, lighter the density,
more plus, darker.
CONTRAST
0
Sets the contrast (contrast between light and dark of the image) of the image.Select from "-7"
- "+7". More negative, lower the contrast, more plus, higher.
SMOOTH
2
Sets the smoothing of the image. Select from "0"~"7".
In the range of "0"~"6", larger the value, smoother the image. However, too large in value
makes the image blur.
Selecting "7" gives the same smoothing result as "0" is selected. The image will be printed in
one to one scale.
Recommended is"2" or "3".
DENSITY MAX CH1, 2 : 300
Sets the max. density of the printed image. Set between "150" ~ "400".
Ch3 : 400
Higher the value, higher the max. density. Lower the value, lower the
CH4 - : 300
max. density.
DENSITY MIN
20
Sets the min. density of the printed image. Set between "10" ~ "30". Higher the value, higher
the min. density. Lower the value, lower the min. density.
ILLUMINATION
2000
Sets the luminance of the light table. Normally, this value is specified and sent from the diag-
nostic device. Set between "0" ~ "38000".
AMBIENT
10
Sets the ambient light. Normally, this value is specified and sent from the diagnostic device.
LIGHT
Set between "0" ~ "38000".
FLIP
OFF
Selects whether to flip the print image. Select from OFF (not reversed) or ON (reversed).
Selecting "ON(reversed)" will print the image laterally reversed.
User Level
1
2
3
Disp
●
●
●
Ope
●
Summary of Contents for DRYPRO Model 793
Page 8: ...About This Manual 6 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 18: ...Ch 1 Safety Precautions Warnings 16 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 28: ...Ch 2 Product Outline 26 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 48: ...Ch 3 Operation 46 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 100: ...Ch 4 Maintenance Mode 98 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 124: ...Ch 5 Web Maintenance Mode 122 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 150: ...Ch 7 Care Maintenance 148 DRYPRO Vstage MODEL 793 Operation Manual Ver 1 01 2005 1 ...
Page 154: ......
Page 155: ......