81
Default Tolerance Setting
Memo
•
The tolerance is set to the following values under the initial settings.
Lower limit: -1.00 Upper limit: 1.00
Operating Procedure
Start the procedure from the <Default data setup> screen.
1
Move the cursor to “Default tolerance”, and then
press the [Confirmation] key.
The <Default tolerance> screen
is displayed.
2
Use [
] or [
] to move the cursor to the item to be
set, and then press the [Confirmation] key.
The <Edit tolerance> screen
is displayed.
3
Use [
]/[
] or [
]/[
] to move the cursor to the
item to set and press the [Confirmation] key to
change the item.
•
When the selected item is not currently checked,
pressing the
[Confirmation] key
causes the item to be
checked, after which the setting value can be changed.
•
Press
[
] or [
]
to move the cursor to the setting value
area. Pressing the
[Confirmation] key
causes the cursor
to appear on the value. Press
[
] or [
]
to change the
value. Move the cursor between the digits of the value by
pressing
[
] or [
]
.
Memo
If “Audi2000(ΔEc)” or “Audi2000(ΔEp)” is selected under
“Sample” - “Display type”, not only is moving between pages
possible with the [
] and [
] keys from the displayed page
when the <Default tolerance> screen is selected, but if
Audi2000(ΔEc) or Audi2000(ΔEp) is enabled, in addition to
the tolerance for each angle, the tolerances for average
(mΔEc or mΔEp) and maximum (ΔEcMax or ΔEpMax) can also
be set. When setting the tolerance for mΔEc, mΔEp, ΔEcMax
or ΔEpMax, setting a value for any angle will cause that value
to be shown for all angles in the tolerance setting screen.
Settings
{
ΔL*, Δa*, Δb*, ΔC*, ΔH*, ΔFF: -20.00 to 20.00
{
Other than the above: 0.00 to 20.00
•
Press the
[Confirmation] key
to confirm every time the
setting of an item is changed.
4
After all settings have been configured, use [
] or
[
] to move the cursor to “OK”, and then press the
[Confirmation] key.
The setting is confirmed and the screen returns to the
<Default tolerance> screen
.
If
[ESC]
is pressed during setup, the settings will not be
changed and the screen will return to the
<Default
tolerance> screen
.
5
Press [ESC] to return to the <Default data setup> screen.
Sample
Setting
0002 TARGET-2
Default data setup
Default tolerance
Parametric coef.
Warning level
80%
Target
Default tolerance
1.00
-1.00
∆FF
∆L*
∆E*ab
∆C*
∆H*
0002 TARGET-2
10°/D65
None
-15°
15°
25°
45°
75°
110°
1.00
0.00
1.00
0.00
1.00
0.00
1.00
0.00
1.00
0.00
1.00
0.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1
2
Target
Default tolerance
1.00
-1.00
∆FF
∆L*
∆E*ab
∆C*
∆H*
0002 TARGET-2
1 10°/D65
2
None
-15°
15°
25°
45°
75°
110°
1.00
0.00
1.00
0.00
1.00
0.00
1.00
0.00
1.00
0.00
1.00
0.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
1.00
-1.00
Edit tolerance
-15°
∆E*ab
∆L*
+01.00
+00.00
+01.00
-00.00
Summary of Contents for CM-M6
Page 1: ...Instruction Manual Please read before using the instrument...
Page 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Page 102: ...99...
Page 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Page 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Page 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Page 126: ...123...
Page 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...