54
3
En
vir
onment Set
tings/Other Set
tings
Pass/Fail
Edit the tolerance that will be used as the judgment criteria, and set the warning level and the parametric
coefficient default value.
If the color difference between the sample and the target exceeds the tolerance, the column of the display color
value relevant to the measurement display will be displayed in red. In addition, if even one display color value
exceeds the tolerance, judgment will be shown as “Fail”.
If the color difference between the sample and the target exceeds the warning level amount, the column of the
display color value relevant to the measurement display will be displayed in yellow. In such cases, even if other
display color values do not exceed the tolerance, judgment will be shown as “Warning”.
Memo
•
The default judgment criteria is set before judgment criteria for each target is set. For details, refer to page 80
“Default Data Settings”.
•
The tolerance setting items are the same as those for the color space or index currently selected for the display
conditions.
•
The optional Color Data Software “SpectraMagic DX” allows for easy configuration and use of judgment criteria.
•
Display the target to be set in advance before moving to <Target menu>.
Operating Procedure
Start the procedure from the <Target menu> screen.
1
Use [
] or [
] to move the cursor to “Pass/Fail”,
and then press the [Confirmation] key. The <Pass/
Fail> screen will be displayed.
Target menu [0002]
Target
Delete data
Pass/Fail
Data management
Edit name
TARGET-2
Target menu [0002]
Pass/Fail
Target
Edit tolerance
Parametric coef.
Warning level
80%
Summary of Contents for CM-M6
Page 1: ...Instruction Manual Please read before using the instrument...
Page 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Page 102: ...99...
Page 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Page 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Page 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Page 126: ...123...
Page 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...