Set the instrument on the zero calibration box while
referring to page 21 “Zero Calibration Box”.
4
Press the measurement key or the measurement
button.
Zero calibration will be performed.
When the zero calibration is completed, a screen
prompting white calibration is displayed. Continue to step
3 on the following page to perform white calibration.
Notes
Do not move the instrument during zero calibration.
Memo
White calibration must be performed when performing zero
calibration.
After zero calibration is complete, a screen prompting
white calibration is displayed. Continue to the next section
to perform white calibration.
Calibration
Please set the
instrument on the zero
calibration box and
press the measurement
button.
Calibration
Calibrating to zero
CALIBRATION
Calibration
Please set the
instrument on the white
calibration plate and
press the measurement
button.
White ID 1000001