79
Direction
Memo
The direction is set to “Double-Path” when the instrument is shipped from the factory.
Operating Procedure
Start the procedure from the <Measurement cond.> - <Display cond.> screen.
1
Use [
] or [
] to move the cursor to “Direction”,
and then press the [Confirmation] key.
The <Direction> screen
is displayed.
2
Use [
] or [
] to move the cursor to the desired
direction.
Selections
{
Double-Path
{
Left
Memo
Data obtained with “Left” selected may have lower
performance than data with “Double-Path” selected.
Moreover, measuring with “Left” selected requires the use of
a sample with a flat measurement surface, and the
instrument must not be tilted when measurement is
performed.
3
Press the [Confirmation] key to confirm, and then
press [ESC].
The selection is confirmed, and the screen returns to the
previous screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed, and the screen will
return to the <Display cond.> screen.
Sample
Measurement cond.
Display cond.
Display type
Equation
CMC
Color space
L*C*h
Direction
Double-Path
Sample
Measurement cond.
Display cond.
Display type
Equation
CMC
Color space
L*C*h
Direction
Double-Path
Direction
Double-Path
Left
Summary of Contents for CM-M6
Page 1: ...Instruction Manual Please read before using the instrument...
Page 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Page 102: ...99...
Page 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Page 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Page 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Page 126: ...123...
Page 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...