71
Manual Average Option
Select the save method when performing manual averaging.
Operating Procedure
Start the procedure from the <Measurement cond.> - <Measurement setup> screen.
1
Use [
] or [
] to move the cursor to “Manual
average option”, and then press the [Confirmation]
key.
The <Manual average option> screen
is displayed.
2
Use [
] or [
] to move the cursor to “Manual save”
or “Auto save”, and then press the [Confirmation]
key.
Selections
{
Manual save
{
Auto save
3
Press the [Confirmation] key to confirm, and then
press [ESC].
Notes
The selection is confirmed and the screen returns to the
previous screen.
Tilt Detection
This instrument features a warning function that alerts users if the instrument is tilted beyond the measurement
range.
Turning this function ON will cause a warning to appear if the instrument is tilted at an angle that cannot be
corrected.
Memo
Tilt detection is set to “ON” under the initial settings.
Operating Procedure
Start the procedure from the <Measurement cond.> - <Measurement setup> screen.
1
Use [
] or [
] to move the cursor to “Tilt
detection”, and then press the [Confirmation] key.
The <Tilt detection> screen
is displayed.
Sample
Measurement cond.
Measurement setup
Measurement angle
Manual average option
Manual save
Manual average
Auto average
Manual average option
Manual save
Auto save
Sample
Measurement cond.
Measurement setup
Tilt detection
OFF
Summary of Contents for CM-M6
Page 1: ...Instruction Manual Please read before using the instrument...
Page 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Page 102: ...99...
Page 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Page 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Page 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Page 126: ...123...
Page 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...