95
Alarm contact
The alarm contact is closed during normal operation
(N/C). It opens in the case of alarm or power outage.
As a result, a failure message is provided even in the
case of power failure or line breakage (fail-safe behav-
ior). For contact ratings, see Specifications.
17
18
Alarm
R3
Error messages can also be signaled by a 22-mA output current (see Error messages
and Configuration Output 1/Output 2).
Operating behavior of the alarm contact: see Operating States table.
The alarm delay time delays the color change of the display backlighting to red,
the 22-mA signal (if configured), and the alarm contact switching.
Configuration
Menu item
Action
Choices
Alarm delay
Enter value using
keys.
Press enter to confirm.
Sensocheck
Select Sensocheck
(continuous monitoring
of glass and reference
electrode)
Select ON or OFF using
keys.
Press enter to confirm.
(At the same time,
Sensoface is activated.
With OFF, Sensoface is
also switched off.)