In the alarm menu you can configure flow monitoring. When you have
set CONTROL to FLOW, you can specify 2 additional limit values for
maximum and minimum flow.
If the measured value lies outside this range, an alarm message and a
22-mA error signal (if configured) will be generated.
Configuration
Menu item
Action
Choices
Select function of
CONTROL input
Select using
keys.
Press enter to confirm.
PARSET
(selecting parameter
set A/B via signal at
CONTROL input)
Flow
(for connecting a pulse-
output flow meter)
Adjust to flow meter
With "Flow" selected,
you must adjust the
device to the flow meter
used.
Enter value using arrow
keys.
Press enter to confirm.
12000 pulses/liter
Display
Flow measurement in measuring mode
Display
Flow measurement (sensor monitor)
5