![Knick Stratos Pro A401B PH User Manual Download Page 140](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199140.webp)
140
Diagnostics
Display
Menu item
Displaying the logbook entries
Select LOGBOOK using
, press enter to confirm.
Using the
keys, you can scroll backwards and
forwards through the logbook (entries -00-...-99-),
-00- being the last entry.
If the display is set to date/time, you can search for a
particular date using the
keys.
Press
to view the corresponding message text.
If the display is set to the message text, you can
search for a particular message using the
keys.
Press
to display the date and time.
Press meas to return to measurement.
Extended logbook / Audit Trail (via TAN)
Using the
keys, you can scroll backwards and
forwards through the extended logbook (entries
-000-...-199-), -000- being the last entry.
Display: CFR
Audit Trail also records function activations
(CAL CONFIG SERVICE), some Sensoface messages
(cal timer, wear) and opening of the enclosure.