Specifications
pH/mV input
Input for pH or ORP sensors or ISFET
Input
Glass electrode or ISFET
Input
Reference electrode
Input
ORP electrode (e.g. platinum) or auxiliary
electrode for impedance measurement
Measuring range
–1500 ... +1500 mV
Display range
pH value
–2.00 ... 16.00
ORP
–1999 ... +1999 mV
Glass electrode input
5)
Input resistance
> 1 x 10
12
Ω
Input current
< 1 x 10
-12
A
Impedance range
0.5 ... 1000 MΩ (± 20%)
Reference electrode input
5)
Input resistance
> 1 x 10
10
Ω
Input current
< 1 x 10
-10
A
Impedance range
2)
0.5 ... 200 kΩ (± 20%)
Measurement error
1,2,3)
pH value
< 0.02
TC: 0.002 pH/K
mV value
< 1 mV
TC: 0.1 mV/K
pH sensor standardization
*
pH calibration
Operating modes
BUF
Calibration with Calimatic
automatic buffer recognition
MAN
Manual calibration with
entry of individual buffer values
DAT
Data entry of pre-measured electrodes
Product calibration
Calimatic buffer sets
*
-01- Mettler-Toledo
2.00/4.01/7.00/9.21
-02- Knick CaliMat
2.00/4.00/7.00/9.00/12.00
-03- Ciba (94)
2.06/4.00/7.00/10.00
-04- NIST technical
1.68/4.00/7.00/10.01/12.46
-05- NIST standard
1.679/4.006/6.865/9.180
-06- HACH
4.01/7.00/10.01
-07- WTW techn. buffers 2.00/4.01/7.00/10.00
-08- Hamilton
4.01/7.00/10.01/12.00
-09- Reagecon
2.00/4.00/7.00/9.00/12.00
-10- DIN 19267
1.09/4.65/6.79/9.23/12.75
-U1-
Specifiable buffer set with 2 buffer solutions
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...