Specifications
Limit values Rel1/Rel2
Rel1/Rel2 contacts, floating, but inter-connected
Contact ratings
AC
< 250 V / < 3 A / < 750 VA
DC
< 30 V / < 3 A / < 90 W
Contact response
*
N/C or N/O
Response delay
*
0000 ... 9999 s
Setpoints
*
As desired within range
Hysteresis
*
User-defined
PID process controller
Output via Rel1/Rel2 relay contacts (see limit values)
Setpoint specification*
Within selected range
Neutral zone *
Max. 50 % of selected range
Proportional action*
Controller gain
Kp: 0010 ... 9999 %
Integral action*
Reset time
tr: 0000 ... 9999 s (0000 s = no integral action)
Derivative action*
Rate time
td: 0000 ... 9999 s (0000 s = no derivative action)
Controller type
*
Pulse length controller or pulse frequency controller
Pulse period
*
0001 ... 0600 s, min. ON time 0.5 s (pulse length controller)
Max. pulse frequency
*
0001 ... 0180 min
–1
(pulse frequency controller)
Real-time clock
Different time and date formats selectable
Power reserve
> 5 days
Display
LC display, 7-segment with icons, colored backlighting
Main display
Character height approx. 22 mm, unit symbols approx. 14 mm
Secondary display
Character height approx. 10 mm
Text line
14 characters, 14 segments
Sensoface
3 status indicators (friendly, neutral, sad face)
Mode Indicators
meas, cal, conf, diag
Further icons for configuration and messages
Alarm indication
Display blinks, red backlighting
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...