77
Configuration
Menu item
Action
Choices
Current range
Select 4-20 mA or 0-20
mA range using
keys.
Press enter to confirm.
Process variable
Select using
keys:
PH: pH value
ORP: Redox potential
TMP: Temperature
Press enter to confirm.
Current start
Modify digit using
keys,
select next digit using
keys.
Press enter to confirm.
Current end
Enter value using
keys.
Press enter to confirm.
Assignment of measured values: Current start and current end
Output current
[pH]
20
4
14
0
20
4
7
5
[mA]
Example 1: Range pH 0...14
[mA]
Output current
7
5
5
Example 2: Range pH 5...7
Advantage: Higher resolution in
range of interest
[pH]