143
Service
Menu item
Remark
Displaying the currently measured values (sensor
monitor) with HOLD mode activated:
Select MONITOR using
, press enter to confirm.
Select variable in the bottom text line using
.
The selected parameter is shown in the main display.
As the device is in HOLD mode, you can perform
validations using simulators without influencing the
signal outputs.
Hold meas depressed for longer than 2 sec to return
to Service menu. Press meas once more to return to
measurement.
SENSOR / TTM
Resetting the adaptive maintenance timer
Here, the interval is reset to its initial value. To do so,
select “TTM RESET = YES“ and confirm by pressing
enter.
SENSOR / AUTOCLAVE
Incrementing the autoclaving counter
After having completed an autoclaving process,
you must increment the autoclaving count.
To do so, select “YES” and confirm by pressing enter.
The device confirms with
“INCREMENT AUTOCLAVE CYCLE“.