131
Display
Action
Remark
The device returns to
measuring mode.
From the blinking
CAL mode indicator
you see that product
calibration has not
been terminated.
Product calibration
step 2
Display (3 sec)
Now the device is in
HOLD mode.
The stored value is
displayed (blinking) and
can be overwritten with
the measured sample
value.
Press enter to proceed.
Display of new asym-
metry potential (based
on 25°C).
Sensoface is active.
To exit calibration:
Select MEAS,
then press enter
To repeat calibra-
tion: Select REPEAT,
then press enter
End of calibration
After end of calibration, the outputs remain in
HOLD mode for a short time.
Product Calibration (pH)