![Knick Stratos Pro A401B PH User Manual Download Page 155](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199155.webp)
155
Specifications
Keypad
Keys: meas, menu, info, 4 cursor keys, enter
HART communication
HART version 6
Digital communication by FSK modulation of output current 1
Device identification, measured values, status and messages,
parameter setting, calibration, records
Conditions
Output current ≥ 3.8 mA and load resistance ≥ 250 Ω
IrDA interface
Infrared interface for firmware update
FDA 21 CFR Part 11
Access control by editable passcodes
Logbook entry and flag via HART in the case of configuration changes
Message and logbook entry when enclosure is opened
Diagnostics functions
Calibration data
Calibration date, zero, slope, response time
Device self-test
Display test, automatic memory test (RAM, FLASH, EEPROM),
module test
Logbook
100 events with date and time
Extended logbook (TAN)
Audit Trail: 200 events with date and time
Service functions
Sensor monitor
Display of direct sensor signals
Current source
Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)
Relay test
Manual control of the four switching contacts
Manual controller
Controller output entered directly (start of control process)
IrDA
Activating the IrDA function
Passcodes
Assigning passcodes for menu access
Factory setting
Resetting all parameters to factory setting
TAN
Enabling optionally available additional functions
Data retention
Parameters, calibration data, logbook > 10 years (EEPROM)
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...