![Knick Stratos Pro A401B PH User Manual Download Page 141](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199141.webp)
141
Diagnostics
Display
Menu item
Display examples:
Display of currently measured values
(sensor monitor)
Select MONITOR using
, press enter to confirm.
Use the
keys to select the desired parameter
from the bottom line of the display: mV_PH mV_ORP
RTD R_GLASS R_REF I-INPUT (for digital sensors also:
OPERATION TIME SENSOR WEAR LIFETIME CIP SIP
AUTOCLAVE, for ISM sensors in addition:
ACT (adaptive calibration timer)
TTM (adaptive maintenance timer)
DLI (Dynamic Life Time Indicator). The selected
parameter is shown in the main display.
Press meas to return to measurement.
Display mV_pH
(for validation, sensor can be immersed in a calibra-
tion solution, for example, or the device is checked by
using a simulator)
Display of remaining dynamic lifetime
(only for digital sensors, however not for MEMOSENS)
Display of sensor operating time
(for digital sensors only)
Version
Display of device type, software/hardware version,
and serial number for all device components.
Use the
keys to switch between software and
hardware version. Press enter to proceed to next
device component.
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...