Specifications
I input (TAN)
Current input 0/4 ... 20 mA / 50 Ω for external temperature signal
Start/end of scale
Configurable –20 ... 200 °C (-4 ... 392 °F)
Characteristic
Linear
Measurement error
1.3)
< 1% current value + 0.1 mA
HOLD input
Galvanically separated (optocoupler)
Function
Switches device to HOLD mode
Switching voltage
0 ... 2 V (AC/DC)
HOLD inactive
10 ... 30 V (AC/DC) HOLD active
CONTROL input
Galvanically separated (optocoupler)
Function
Selecting parameter set A/B or flow measurement
Parameter set A/B
Control input
0 ... 2 V (AC/DC)
10 ... 30 V (AC/DC)
Parameter set A
Parameter set B
FLOW
Pulse input for flow measurement 0 ... 100 pulses/s
Pulse amplitude 10 ... 30 V DC
Message
via 22 mA, alarm contact or limit contacts
Display
00.0 … 99.9 l/h
Output 1
0/4 ... 20 mA, max. 10 V, floating (galv. connected to output 2)
Process variable *
pH, ORP or temperature
Characteristic
Linear
Overrange *
22 mA in the case of error messages
Output filter *
PT
1
filter, time constant 0 ... 120 s
Measurement error
1)
< 0.25 % of current value + 0.025 mA
Start/end of scale
*
Configurable within selected range
Minimum span
pH 2.00 / 200 mV / 20 K / 36 °F
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...