Pulse Length / Pulse Frequency Controller
Pulse length controller (PLC)
The pulse length controller is used to operate a valve as an actuator.
It switches the contact on for a time that depends on the controller
output. The period (pulse length) is constant. A minimum ON time of
0.5 sec is maintained even if the controller output takes correspond-
ing values (Y=0: Off).
Output signal (relay contact) of pulse length controller
ON time (Y = 80 %)
ON time (Y = 20 %)
Pulse length
1
0
1
0
Pulse frequency controller (PFC)
The pulse frequency controller is used to operate a frequency-
controlled actuator (metering pump). It varies the frequency with
which the contacts are switched on. The maximum pulse frequency
[pulses/min] can be defined. It depends on the actuator.
The contact ON time is constant. It is automatically calculated from
the user-defined maximum pulse frequency:
Output signal (relay contact) of pulse frequency controller
ON time
Pulse frequency (Y = 20%)
Pulse frequency (Y = 80%)
1
0
1
0
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...