Configuration
Menu item
Action
Choices
Calibration timer
Adjust CALTIMER using
keys:
OFF: No timer
ADAPT: Maximum cal
cycle (adjust in the next
step)
FIX: Fixed cal cycle (adjust
in the next step)
Press enter to confirm.
With ADAPT, the calibra-
tion cycle is automatically
reduced depending on
the sensor load (high
temperatures and pH
values) and for digital
sensors also depending
on the sensor wear
Calibration cycle
Only with FIX/ADAPT:
Modify digit using
keys,
select next digit using
keys.
Press enter to confirm.
5
Note for the calibration timer:
When Sensocheck has been activated in the Configuration – Alarm
menu, the expiration of the calibration interval is indicated by
Sensoface:
Display
Status
+
+
Over 80 % of the calibration interval has
already past.
The calibration interval has been exceeded.
The calibration timer settings apply to both parameter sets A and B.
The time remaining until the next due calibration can be seen in the
diagnostics menu (see Diagnostics chapter).
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...