Specifications
Output 2
0/4 ... 20 mA, max. 10 V, floating (galv. connected to output 1)
Process variable *
pH, ORP (mV) or temperature
Characteristic
Linear
Overrange *
22 mA in the case of error messages
Output filter *
PT
1
filter, time constant 0 ... 120 s
Measurement error
1)
< 0.25 % of current value + 0.025 mA
Start/end of scale
*
Configurable within selected range
Minimum span
pH 2.00 / 200 mV / 20 K / 36 °F
Alarm contact
Relay contact, floating
Contact ratings
AC
< 250 V / < 3 A / < 750 VA
DC
< 30 V / < 3 A / < 90 W
Contact response
N/C (fail-safe type)
Response delay
0000 ... 0600 s
Wash contact
Relay contact, floating
For controlling a cleaning system
Contact ratings
AC
< 250 V / < 3 A / < 750 VA
DC
< 30 V / < 3 A / < 90 W
Contact response
*
N/C or N/O
Interval
*
000.0 ... 999.9 h (000.0 h = cleaning function switched off)
Cleaning duration
*
0000 ... 1999 s
or
Parameter set A/B
For signaling parameter set A/B
Contact ratings
AC
< 250 V / < 3 A / < 750 VA
DC
< 30 V / < 3 A / < 90 W
Contact response
*
Contact open:
Parameter set A active
Contact closed:
Parameter set B active
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...