![Knick Stratos Pro A401B PH User Manual Download Page 118](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199118.webp)
118
Digital Sensors
Connecting a Digital Sensor
Step
Action/Display
Remark
Connect sensor
Before a digital sensor
is connected, the error
message “No sensor“ is
displayed.
Wait until the
sensor data are
displayed.
MEMO SENS
The hourglass in the
display blinks.
Check sensor
data
MEMO SENS
View sensor information
using
keys,
press enter to confirm.
Display color changes to
green.
Sensoface is friendly
when the sensor data are
okay.
Go to
measuring
mode
Press meas, info or enter
After 60 sec the device
automatically returns to
measuring mode (time-
out).
Replacing a Sensor
A sensor should only be replaced during HOLD mode to prevent un-
intended reactions of the outputs or contacts. When you first want to
calibrate the new sensor, it can also be replaced in calibration mode.