81
Connecting a Digital Sensor
Step
Action/Display
Remark
Connect sensor
Before a digital sensor
is connected, the error
message “No sensor“ is
displayed.
Wait until the sensor
data are displayed.
The hourglass in the
display blinks.
Check sensor data
View sensor informa-
tion using
keys,
confi rm with
enter
.
The display color changes
to
green
.
Sensoface is happy.
Go to measuring
mode
Press
meas
,
info
, or
enter
After 60 sec the device
automatically returns to
measuring mode
(timeout).
Possible error messages
Sensor devaluated
Replace sensor
When this error message
appears, the sensor can-
not be used any more.
Sensoface is sad.
Sensor defective
Replace sensor
When this error message
appears, the sensor
cannot be used.
Sensoface is sad.