70
Controller Functions
Proportional action Y
P
Setpoint - Meas. value
Constant
Y
P
=
*
Controller output Y =
I action
with:
Y
P
Proportional
action
T
R
Reset
time
[s]
T
D
Rate
time
[s]
K
C
Controller
gain
[%]
Constant
5 (for pH)
500
mV
(for
ORP)
D action
dY
P
dt
1
T
R
P action
Controller Equations
Y
P
+
Y
P
dt
+ T
D
K
C
Neutral Zone
Tolerated deviation from desired value.
With the setting “1 pH”, for example, a deviation of ± 0.5 pH from the
desired value is tolerated.
100 %
50 %
X
w
Process variables: pH/ (mV),
underneath: temp [K]
Contr
oller
output Y
K
C
= 500 %
K
C
= 200 %
K
C
= 100 %
K
C
= 50 %
1 / (100)
2 / (200)
3 / (300)
4 / (400)
Proportional Action (Gradient K
C
[%])
Deviation
5 / (500)
[10]
[20]
[30]
[40]
[50]