82
Digital Sensors
Sensor Replacement
A digital sensor should only be replaced during HOLD mode to pre-
vent unintended reactions of the outputs or contacts. When you fi rst
want to calibrate the new sensor, it can also be replaced in calibration
mode.
Step
Action/Display Remark
Select HOLD mode
Press any key to call
the selection menu,
select HOLD using
the
keys, press
enter
to confi rm.
Now the device is in
HOLD mode. The HOLD
mode can also be acti-
vated externally via the
HOLD input.
During HOLD the output
current is frozen at its
last value or set to a fixed
value.
Disconnect and
remove old sensor
Install and connect
new sensor.
Temporary messages
which are activated dur-
ing the replacement are
indicated but not output
to the alarm contact and
not entered in the log-
book.
Wait until the sensor
data are displayed.